ELx808
Absorbance Reader
The ELx808™ Absorbance Microplate Reader offers proven performance for many applications, from endpoint to fast kinetic protocols. The ELx808 is widely recognized as offering the industry-leading 4-Zone™ temperature control, which optimizes conditions for even the most temperature sensitive assays. Its high quality optical design ensures excellent precision, accuracy and linearity. Gen5 Software controls the ELx808 for easy data capture and powerful analysis.
- Superior 4-Zone™ incubation
- Fast kinetics, plus endpoint and linear well scanning
- Optical design eliminates crosstalk
- Extensive data analysis with available Gen5 Microplate Reader and Imager Software
APPLICATIONS
The ELx808 with 4-Zone™ Incubation and fast reading speed is well suited to temperature sensitive kinetic assays and many other colorimetric and turbidimetric protocols.
FEATURES
4-Zone incubation
Unique heated track and carrier design minimizes evaporation and edge effect. Proven design is optimal for temperature sensitive applications like endotoxin analysis.
Fast kinetics, plus endpoint and linear well scanning
With measurements as fast as 6 seconds, ELx808 is ideal for fast kinetic assays. Linear well scanning collects multiple data points across the well for applications such as agglutination studies.
Optical design eliminates crosstalk
The multi-channel, staggered optics design of the ELx808 virtually eliminates channel-to-channel crosstalk.
Extensive data analysis with available Gen5 Microplate Reader and Imager Software
For data collection and powerful quantitative and qualitative analysis, the available Gen5 software offers intuitive protocol creation and flexible reporting and exporting functions.
GENERAL
Detection modes | Absorbance |
Read methods | End point, kinetic, linear scanning |
Microplate types | 96-well plates |
Temperature control | 4-Zone incubation to 50 °C +0.2 °C at 37 °C |
Shaking | Linear |
Software | Gen5 Reader Control Software Gen5 Secure for 21 CFR Part 11 compliance (option) |
ABSORBANCE
Light source | Tungsten halogen |
Detector | photodiode |
Wavelength selection | filters |
Wavelength range | 380 to 900 nm 340 to 900 nm (ELx808IU) |
Dynamic range | 0 to 4.0 OD |
Resolution | 0.001 OD |
Pathlength correction | no |
Filter wheel capacity | 6 positions |
Filters supplied | 4 filters (5 with UV option) |
OD accuracy | < 1% at 2.5 OD |
OD linearity | <1% at 2.5 OD < 2% at 3.5 OD |
OD repeatability | < 0.5% at 2.5 OD < 1.5% at 3.5 OD |
Reading speed (kinetic) | 96 wells: 8 seconds |
PHYSICAL CHARACTERISTICS
Power | 100 – 240 Volts AC. 50/60 Hz |
Dimensions | 15.5″W x 16″D x 8.75″H |
Weight | 30 lbs (13.6 kg) |
REGULATORY
Regulatory | CE and TUV marked. RoHS Compliant. Models for In Vitro Diagnostic use are available. |
Part #
|
ELX808 | ELX808IU |
---|---|---|
96-well
|
• | • |
380-900 nm
|
• | |
340-900 nm
|
• | |
Incubation
|
• | |
Shaking
|
• | • |
Software
|
Includes Gen5 Reader Control Software | Includes Gen5 Reader Control Software |
Reviews
There are no reviews yet.