Cytation™ 5 combines automated digital microscopy and conventional microplate detection in a configurable, upgradable platform. The microscopy module offers up to 60x magnification in fluorescence, brightfield, high contrast brightfield, color brightfield and phase contrast for maximum applications reach. The multi-mode detection modules include filter- and monochromator-based fluorescence detection, luminescence and UV-Vis absorbance detection. Gen5™ software provides complete control over all imaging and data capture, plus powerful image and data analysis.
- Modular digital microscopy and multi-mode detection, each with independent optical components
- CO2/O2 control, incubation and shaking for live cell assays
- Wide field of view camera for fast whole well imaging
- Label-free cell counting for cell proliferation studies
- More than 20 imaging filter/LED cube colors reach many applications
- Patented Hybrid optics: sensitivity of filters, flexibility of monochromator
- Variable bandwidth monochromators optimize fluorophore detection
- 100 mW Alpha Laser: Speed and performance-based excitation for Alpha assays
- Modular design
- Peltier Cooling Module maintains environmental stability for uncompromised assay integrity
Seahorse XF Imaging and Normalization System
Cytation 5 is a key component of Agilent Technologies’ Seahorse XF Imaging and Normalization System. XFe Analyzers simultaneously measure the two major cellular energy-producing pathways, mitochondrial respiration and glycolysis, in live cells and in real time. The integrated solution with Cytation 5 combines cellular analysis and imaging for an intuitive workflow to streamline XF data interpretation.
WHAT’S NEW
The Scratch Assay Starter Kit automates cell migration and wound healing assays, used along with BioTek’s Lionheart FX Automated Microscope or Cytation 5 Cell Imaging Multi-Mode Reader.
APPLICATIONS
- Cell migration and invasion assays
- Live cell imaging
- Multi-mode reader applications
- Immunofluorescence
- Phenotypic assays
- Histology
The Scratch Assay Starter Kit automates cell migration and wound healing assays:
FEATURES
Digital microscopy and multi-mode detection
Cytation 5’s patented design enables automated digital microscopy to be combined with conventional multi-mode detection to suit a broad range of imaging and biochemical workflows. Along with Gen5 software, Cytation’s imaging module comprises Augmented Microscopy – the ability to capture>process>analyze>publish high-quality data and images, to obtain both qualitative and quantitative information on live and fixed cell imaging applications.
Cytation 5 is well suited for live cell assays, with temperature control to 65 °C, plus linear and orbital shaking. A gas controller module is available to help regulate CO2 and O2 levels in the system for longer-term live cell requirements. A dual reagent injector provides quick inject/image capability for observing fast reactions.
Wide field of view camera for fast whole well imaging
Capture an image of an entire well with the wide FOV camera – the image at left shows an image captured in a single well of 384-well plate at 4x. Further benefits of the wide FOV camera are fast capture speeds for large samples and statistically significant sampling for cellular population analysis.
Label-free cell counting for cell proliferation studies
It’s fast and simple to count live cells using the available high contrast brightfield option. Higher contrast allows a direct and label-free method for object identification using Gen5 software to count cells and characterize cell proliferation in kinetic studies.
Patented Hybrid optics for fluorescence detection
Cytation 5’s hybrid optical design offers filter-based fluorescence optics for high transmission and high sensitivity, while the available monochromator optics offers convenience and assay flexibility. Multiple fluorescence measurement modes include fluorescence intensity, time-resolved fluorescence and fluorescence polarization.
Variable bandwidth monochromators
The bandwidths of the monochromators in Cytation 5’s fluorescence detection module are selectable from 9 nm to 50 nm, in 1 nm increments. Variable bandwidths allow you to optimize detection conditions for fluorophores that have unique excitation or emission parameters, for example, a narrow Stokes shift, to provide better performance than conventional monochromators.
100 mW Alpha Laser
Laser-based excitation offers the power required to get great sensitivity and fast reading speeds, which are critical for Alpha assays.
Modular, upgradable design
Cytation 5 is a modular system, with imaging optics and multi-mode optics available in a variety of configurations and upgrade paths. The separate modules ensure excellent performance in imaging and all detection modes, without compromise Choose imaging only, with available phase contrast. Or, add filter- and/or- monochromator-based fluorescence and luminescence, UV-Vis absorbance along with available laser-based Alpha detection.
Peltier Cooling for environmental stability
The Peltier Cooling Module promotes a rapid interior cool down after incubated processes, to allow efficient switching between multiple applications without unwanted temperature influences. For temperature sensitive assays typically run at ambient, or room temperature, the Cooling Module maintains environmental stability within Cytation, allowing < 1 °C rise in ambient temperature, regardless of external and internal temperature fluctuation.
GENERAL
Detection modes | UV-Vis absorbance Fluorescence intensity Luminescence Fluorescence polarization Time-resolved fluorescence Alpha |
Read methods | Endpoint, kinetic, spectral scanning, well area scanning |
Microplate types | Monochromator: 6- to 384-well plates Filters: 6- to 1536-well plates Imaging: 6- to 1536-well plates |
Other labware supported | Microscope slides, Petri and cell culture dishes, cell culture flasks (T25), counting chambers (hemocytometer) Take3 Micro-Volume Plates |
Temperature control | 4-Zone incubation to 65 °C with Condensation Control; +0.2 °C at 37 °C |
Cooling | Optional Peltier Cooling Module maintains internal temperature with < 1 °C rise over ambient. Provides rapid cooling after incubated processes. |
Shaking | Linear, orbital, double orbital |
Software | Gen5™ Microplate Reader and Imager Software Gen5 Image+ and Image Prime software available for full image analysis Gen5 Secure for 21 CFR Part 11 compliance (option) |
Automation | BioStack and 3rd party automation compatible BioSpa 8 Automated Incubator compatible |
CO2 and O2 control (option) | Range: 0 – 20% (CO2); 1 – 19% (O2) Models for both CO2 and O2 or CO2 only are available |
IMAGING SYSTEM
Imaging mode | Fluorescence, brightfield, high contrast brightfield, color brightfield and phase contrast |
Imaging method | Single color, multi-color, montage, time lapse, z-stacking |
Image processing | Z-projection, digital phase contrast, stitching |
Camera | 16-bit Sony CMOS, standard or wide FOV |
Objective capacity | 6 user-replaceable objectives |
Objectives available | 1.25x, 2.5x (2.25x eff),2.5x (2.75x eff), 4x,10x, 20x, 40x, 60x |
Phase objectives available | 4x, 10x, 20x, 40x |
Image filter cube capacity | 4 user-replaceable fluorescence cubes plus brightfield channel |
Imaging filter cubes available | DAPI, CFP, GFP, YFP, RFP, Texas Red, CY5, CY7,Acridine Orange (ACR OR), CFP-YFP FRET, propidium,Iodide, chlorophyll, phycoerythrin, CY5.5, TagBFP, Alexa568, Ex377 / Em647 |
Imaging LED cubes available | 365 nm, 390 nm, 465 nm, 505 nm, 523 nm, 590 nm, 623 nm, 655 nm, 740 nm |
Automated functions | Autofocus, auto LED intensity, auto exposure |
Autofocus method | Image-based autofocus User-trained autofocus Laser autofocus (option) |
Positional controls | Software control Joystick control (option) |
Image collection rate | Image-based autofocus: 96 wells, 1 color (DAPI), 4x, 6 minutes 96 wells, 3 colors, 4x, 12 minutes Laser autofocus: |
Image Analysis Software option | Gen5 Image+: Advanced image analysis Gen5 Image Prime: Advanced image analysis Gen5 Secure Image+: Advanced image analysis, 21 CFR Part 11 features |
FLUORESCENCE INTENSITY
Light source | Xenon flash |
Detector | PMT for monochromator system PMT for filter system |
Wavelength selection | Quad monochromators (top/bottom) Filters (top) |
Wavelength range | Monochromators: 250 – 700 nm (850 nm option) Filters: 200 – 700 nm (850 nm option) |
Monochromator bandwidth | Variable, from 9 nm to 50 nm in 1 nm increments |
Dynamic range | 7 decades |
Sensitivity | Filters: Fluorescein 0.25 pM (0.025 fmol/well, 384-well plate) Quad Monochromator: |
Reading speed (kinetic) | 96 wells: 11 seconds 384 wells: 22 seconds |
LUMINESCENCE
Wavelength range | 300 – 700 nm |
Dynamic range | >6 decades |
Sensitivity | Monos: 20 amol ATP (flash) Filters: 10 amol ATP (flash), 100 amol (glow) |
FLUORESCENCE POLARIZATION
Light source | Xenon flash |
Detector | PMT |
Wavelength selection | Filters |
Wavelength range | 280 – 700 nm (850 nm option) |
Sensitivity | 1.2 mP standard deviation at 1 nM fluorescein |
TIME-RESOLVED FLUORESCENCE
Light source | Xenon flash |
Detector | PMT |
Wavelength selection | Quad monochromators (secondary mode) Filters (top) |
Wavelength range | Filters: 200 – 700 nm (850 nm option) |
Sensitivity | Filters: Europium 40 fM (4 amol/well, 384-well plate) Monos: Europium 1200 fM (120 amol/well, 384-well plate) |
ABSORBANCE
Light source | Xenon flash |
Detector | photodiode |
Wavelength selection | Monochromator |
Wavelength range | 230 – 999 nm, 1 nm increment |
Monochromator bandwidth | 4 nm (230 – 285 nm), 8 nm (>285 nm) |
Dynamic range | 0 – 4.0 OD |
Resolution | 0.0001 OD |
Pathlength correction | yes |
Monochromator wavelength accuracy | +2 nm |
Monochromator wavelength repeatability | +0.2 nm |
OD accuracy | <1% at 2.0 OD <3% at 3.0 OD |
OD linearity | <1% from 0 to 3.0 OD |
OD repeatability | <0.5% at 2.0 OD |
Stray light | 0.03% at 230 nm |
Reading speed (kinetic) | 96 wells: 11 seconds 384 wells: 22 seconds |
ALPHA DETECTION
Light source | 100 mW 680 nm laser |
Detector | PMT |
Wavelength selection | Filters (top) |
Sensitivity | 100 amol LCK peptide (384-well low volume plate) |
REAGENT INJECTORS (OPTION)
Supported detection modes | All modes |
Number | 2 syringe pumps |
Supported labware | 6- to 384-well plates, Petri and cell culture dishes |
Dead volume | 1.1 mL with back flush |
Dispense volume | 5 – 1000 µL in 1 µL increments |
Plate geometry | 6- to 384-well microplates, Petri dishes |
Dispense accuracy | +1 µL or 2% |
Dispense precision | <2% at 50 – 200 µL |
PHYSICAL CHARACTERISTICS
Power | 250 Watts maximum consumption. |
Dimensions | 20″ D x 16.5″ W x 17.5″ H (50.8 cm x 41.91 cm x 44.5 cm) |
Weight | 80 lbs (36.3 kg) |
REGULATORY
Regulatory | CE and TUV marked. RoHS Compliant. Models for In Vitro Diagnostic use are available. |
Imaging Plus Multi-Mode Configurations
Part # |
CYT5FV
|
CYT5FW |
CYT5FAV
|
CYTFAW |
CYT5MV
|
CYT5MW |
CYT5MPV
|
CYT5MPW |
CYT5MFV
|
CYT5MFW | CYT5MFAV | CYT5MFAW |
---|---|---|---|---|---|---|---|---|---|---|---|---|
Fluorescence imaging
|
•
|
•
|
•
|
• |
•
|
• |
•
|
• |
•
|
• |
•
|
• |
Brightfield imaging
|
•
|
•
|
•
|
• |
•
|
• |
•
|
• |
•
|
• |
•
|
• |
Color brightfield imaging
|
•
|
•
|
•
|
• |
•
|
• |
•
|
• |
•
|
• |
•
|
• |
Phase contrast imaging
|
•
|
•
|
||||||||||
Wide FOV camera
|
•
|
•
|
•
|
•
|
•
|
•
|
||||||
Filter-based fluorescence intensity
|
•
|
•
|
•
|
•
|
•
|
•
|
•
|
•
|
||||
Filter-based fluorescence polarization
|
•
|
•
|
•
|
•
|
•
|
•
|
•
|
•
|
||||
Filter-based time resolved fluorescence
|
•
|
•
|
•
|
•
|
•
|
•
|
•
|
•
|
||||
Filter-based luminescence
|
•
|
•
|
•
|
•
|
•
|
•
|
•
|
•
|
||||
Filter-based Alpha detection
|
•
|
•
|
•
|
•
|
||||||||
Monochromator-based UV-Vis absorbance
|
•
|
•
|
•
|
•
|
•
|
•
|
•
|
•
|
||||
Monochromator-based fluorescence intensity
|
•
|
•
|
•
|
•
|
•
|
•
|
•
|
•
|
||||
Monochromator-based time resolved fluorescence (secondary)
|
•
|
•
|
•
|
•
|
•
|
•
|
•
|
•
|
||||
Monochromator-based Luminescence
|
•
|
•
|
•
|
•
|
•
|
•
|
•
|
•
|
||||
Take3 Plate compatible
|
•
|
•
|
•
|
•
|
•
|
•
|
•
|
•
|
Imaging Only and Multi-Mode Only Configurations
Part # |
CYT5V
|
CYT5W |
CYT5PV
|
CYT5PW | CYT5MF | CYT5MFA | CYT5F | CYT5FA | CYT5M |
---|---|---|---|---|---|---|---|---|---|
Fluorescence imaging
|
•
|
• |
•
|
• | |||||
Brightfield imaging
|
•
|
• |
•
|
• | |||||
Color brightfield imaging
|
•
|
• |
•
|
• | |||||
Phase contrast imaging
|
•
|
• | |||||||
Wide FOV camera
|
•
|
•
|
|||||||
Filter-based fluorescence intensity
|
•
|
•
|
•
|
•
|
|||||
Filter-based
fluorescence polarization
|
•
|
•
|
•
|
•
|
|||||
Filter-based time resolved fluorescence
|
•
|
•
|
•
|
•
|
|||||
Filter-based luminescence
|
•
|
•
|
•
|
•
|
|||||
Filter-based Alpha detection
|
•
|
•
|
|||||||
Monochromator-based UV-Vis absorbance
|
•
|
•
|
•
|
||||||
Monochromator-based Fluorescence intensity
|
•
|
•
|
•
|
||||||
Monochromator-based Time resolved fluorescence (secondary)
|
•
|
•
|
•
|
||||||
Monochromator-based Luminescence
|
•
|
•
|
•
|
||||||
Take3 Plate compatible
|
•
|
•
|
•
|
All configurations include incubation to 65 °C and shaking, and are compatible with BioTek’s Dual Reagent Injection Module, CO2/O2 Gas Controller, and BioStack Microplate Stacker, and BioSpa Automated Incubator.
Reviews
There are no reviews yet.